 ===== Status ===== ===== Status =====
 +
 +== 12 August 2014 ==
 +**Inactive** - Leak detected between bell jar and base plate.  Awaiting parts for the diffusion pump cooling lines flow switch.  Researching replacements for malfunctioning crystal monitor controllers.
 +
 +
 == 05 July 2012 == == 05 July 2012 ==
 Two of the three crystal monitors are currently operational. Two of the three crystal monitors are currently operational.
Line 22: Line 27:
 ===== Maintenance Log===== ===== Maintenance Log=====
  
-07/13/2012 - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days.  - RW+== 09/02/2015 - JH == 
 + 
 +- Chiller water turned black due to the break down of glycol in the cooling water (causing the cooling water to become corrosive).  Chiller is being flushed, sanitized, and refilled by [[http://www.chemaqua.com/index2.asp|Chem-Aqua]]. 
 + 
 +- Update to the waterlines for cooling the evaporation terminals and crystal monitors is complete. 
 + 
 +- Water leak found for line going into diffusion pump, possibly caused by the corrosive water in water the lines. 
 + 
 +== 06/19/2014 - JH == 
 +- Diffusion pump cleaned, and oil changed. 
 + 
 +== 06/19/2013 - RW == 
 + 
 +- Roughing pump oil changed. Old oil in good condition. 
 + 
 +== 07/13/2012 - RW == 
 + 
 +- Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. 
 + 
 +== 07/03/2012 - RW == 
 +  
 +- Roughing pump oil changed. Good color. Some debris. 
 + 
 +== 8/09/2011 - RW ==
  
-07/03/2012 - Roughing pump oil changed. Good color. Some debris. - RW+- Roughing pump oil changed. Old oil was brown and had significant amounts of debris floating in it. Added a permanent maintenance log to the blue binder holding all of the user manuals for the system.
  
-8/09/2011 - Roughing pump oil changed. Old oil was brown and had significant amounts of debris floating in it. Added a permanent maintenance log to the blue binder holding all of the user manuals for the system. - RW +== 6/17/2011 - RW ==
  
-6/17/2011 - Inside of bell jar and all surfaces cleaned. Shutter operation fixed. Ion gauge degaussed; now reads more realistic pressure. - RW+- Inside of bell jar and all surfaces cleaned. Shutter operation fixed. Ion gauge degaussed; now reads more realistic pressure.
  
 ====== Schrader Thermal Evaporator Wngr 489 ====== ====== Schrader Thermal Evaporator Wngr 489 ======
 Stainless steel box evaporator commissioned 1991.  Conflat-sealed except for o-ring-sealed front door.  Base pressure few x 10^-8 torr. Turbo pumped.  Roughing pump is rotary vane 27 cfm pump (?). 3 thermal sources monitored by quartz crystals. Reactive environment possible. Stainless steel box evaporator commissioned 1991.  Conflat-sealed except for o-ring-sealed front door.  Base pressure few x 10^-8 torr. Turbo pumped.  Roughing pump is rotary vane 27 cfm pump (?). 3 thermal sources monitored by quartz crystals. Reactive environment possible.
  
-===== Current Status (June 2013)=====+===== Status ===== 
 +== June 2013 == 
 Upgraded over 2012/13 academic year by Casey Hines. Casey's thesis has a description. {{:theses:hines_bs_2013.pdf|Practical Implementation of a Physical Vapor Deposition System in a Research Environment}} Upgraded over 2012/13 academic year by Casey Hines. Casey's thesis has a description. {{:theses:hines_bs_2013.pdf|Practical Implementation of a Physical Vapor Deposition System in a Research Environment}}
  